Caution should be used for patients with suspected or diagnosed epilepsy.
•
Electrodes should not be placed in direct contact or in close proximity (one inch or less) of each other during
treatment. Electrodes placed in contact or in close proximity can lead to high energy density and skin burns
under or between the electrodes.
•
Care should be used when removing electrodes after treatment, in order to minimize the potential for skin
tearing. Skin should be inspected after removal of electrodes for any signs of tearing or irritation.
•
Do not connect the stimulator to any electrical equipment for combination therapy except the Omnisound
®
family of ultrasounds.
•
ACP does not recommend the use of the Denervated Muscle or Low Intensity DC protocols, and does not
supply the DC or Microcurrent adapter required for use with these protocols. If this device is used with a DC or
Microcurrent adapter, follow the specific treatment guidelines, indications, contraindication, warning and
precautions specified in the DC or Microcurrent adapter manual. Do not use any protocols requiring the use of
the DC or Microcurrent adapter without first having plugged the adapter into the appropriate channel.