OMNISTIM
®
500A USER MANUAL
31
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
2.
Treatment Programs
The MFAC mode contains six pre-programmed modes and a custom program which may be programmed and saved
by the user. The custom program will remain as long as the batteries remain charged or are not removed. All
parameters other than the Carrier frequency may be varied during operation. The following programs are available
in MFAC mode:
NMES EDEMA - MUSCLE PUMP:
Activation of muscle causing contraction of the agonist and antagonist to move the joint. The muscle pump action
compresses fluids into the venous and lymphatic return, while improving blood flow to the edematous tissue.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
35Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
4 seconds
0 – 20 Seconds
Rate scan:
ON
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Channel B delay:
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
4 seconds
1 – 30 seconds
Time off:
12 seconds
0 – 199 seconds
Channels A/B:
ALTERNATE
ALTERNATE or
SIMULTANEOUS
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA
MAFC WAVEFORM