Operational Sequence
1.
Press the ON/OFF button to turn on the power. The display panel will illuminate. There will be no
stimulation current and the unit will be in the SELECT MODALITY/PROGRAM mode.
2.
At this time select the treatment modality desired and the appropriate treatment protocol by using the
modality buttons (2) to select HVPC, LVPC, MFAC or IFC protocols.
3.
Now position the electrodes on the patient, connect the electrode plug(s) and press the START/STOP
button (3/5). The program parameters will disappear and be replaced with the screen seen above.
The bar graph on the lower right of the display shows the relative outputs of the two channels primarily
with respect to Time On and Time Off and Vector operation. The numeric displays on the lower left show
the output intensities into a 500-ohm load. Each time a treatment is commenced, the output current level is
reset to zero; this ensures that your patient cannot receive an initial shock due to the current output,
inadvertently, being left at a high setting from a previous treatment. The patient safety switch is a remote
duplicate of the Start/Stop switch and therefore resets the outputs to zero.
4.
Adjust output by pressing the up/down (4) for channel A and channel B. Once desired output intensity is
selected, press START/STOP button (3/5) to display timer and to initiate treatment.
5.
Press the SELECT PARAMETERS button to modify any settings. Press the MODIFY TIME / PARAM up or
down buttons to adjust the desired treatment time, or other settings, if you wish to change from the default
settings. Set the timer for the desired treatment time. Pressing the switch up or down, increases or decreases the
duration from a minimum of 1 minute to a maximum of 60 minutes. The timer will count down to zero during
treatment; when zero is reached a warning tone is emitted from the unit, the treatment current is switched off and
Treatment Off is displayed. Pressing the START/STOP button (3/5) will shut off the warning tone. If the unit is
left unattended, the warning tone is emitted for ten seconds before power is automatically switched off. Note that
the treatment time is preset for each program but may be adjusted at will during the treatment.
NOTE:
When selecting the MFAC or LVPC alternate or delayed modes the system will allow first set up of
output A followed by set up of output B. Pushing the START/STOP SET output again will start the unit
timing cycles if applicable in the program.
6.
During operation the parameters may be altered within the program at will by selecting the set mode. See
set mode operation and adjustable parameters for each modality.
7.
Once the treatment is completed, remove the electrodes from the patient and disconnect the lead wires from
the electrodes.
8.
Turn the unit off by pressing the power button.
SET OUTPUT A/B
A:00 mA B:00 mA r2.0