Battery Charger / Power Supply Operation
Before using the
OMNISTIM
®
500 battery charger/power supply:
1.
Verify that rechargeable batteries are installed within the
OMNISTIM
®
500 battery compartment.
2.
Check to see that the battery type selector switch located inside the battery compartment has been switched
to the rechargeable position.
If you plug in the charger, after two seconds, the charge LED on the
OMNISTIM
®
500 is illuminated. After the
batteries are fully charged, the LED blinks. If you plug in the charger during treatment or when the unit is ON the
LED turns off. (The charging continues but at a trickle and does not cause the LED to glow.) When you turn OFF
the unit after two seconds the LED comes on to announce charging.
NOTE:
If non-rechargeable batteries are used:
1.
Do not use the charger.
2.
Make sure that the rechargeable/non-rechargeable selector switch in the bottom of the battery
compartment is in the non-rechargeable position.
Life expectancies of batteries under nominal (20 mA IFC, 40 mA MFAC, 100v HVPC) load conditions:
•
4.4Ah NiCad rechargeable batteries - 10 Hours.
•
Alkaline non-rechargeable batteries - 100 Hours.