40 OMNISTIM
®
500A USER MANUAL
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
2. Adjustable Parameters
Pushing the set button during operation brings up the available adjustable parameters. These may be altered during
system operation. To adjust any parameter simply push the parameter set button at any time during operation. This
will display a parameter that may be changed. Repeated pushing of the parameter set button will continue to
display parameters one at a time until all parameters have been displayed. To change the parameter use the
timer/parameter set rocker switch on the front panel to increase, decrease or change the value of the parameter.
Adjustable Parameters
PULSE RATE:
Adjustable from .1 - 125 Hz
INTERPHASE INTERVAL:
Adjustable from 10 to 250 microseconds.
ON - OFF RAMPS:
Adjustable from 0 to 9.9 sec
ON - OFF TIME:
On time adjustable from 0-30 seconds, off time adjustable from 0-199 seconds
AUTO INTENSITY:
Adjusts the output during treatment automatically from 0 to 20% user programmable.
HVPC OUTPUT
PULSE
RISE TIME
>500
NANOSECONDS
INTERPHASE
INTERVAL
100
MICROSECONDS
0-250
VOLTS
HIGH VOLTAGE PULSED CURRENT WAVEFORM:
Continuous or surged operation with fully adjustable ON and
OFF times and ramps allow applications of HVPC therapy.
Reduction of edema, pain control, NMES and soft tissue
management programs can be set up with the
OMNISTIM
®
500’s high volt pulsed current mode.