OMNISTIM
®
500A USER MANUAL
35
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
Low Volt Pulsed Current (LVPC)
1. Treatment Programs
The LVPC mode contains seven pre-programmed modes and a custom program, which may be programmed and
saved by the user. The custom program will remain as long as the batteries remain charged or are not removed.
All parameters may be varied during operation. The following programs are available in LVPC mode:
ANALGESIA SENSORY:
Sensory stimulation activates A-beta fibers causing the release of spinal Enkephalin and Dynorphin to block pain at
the segmental level (Gate Control). Duration of relief is typically from 30 minutes to 2 hours. Fast onset of relief
achieved usually within 20 minutes.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Phase duration:
80 µS
30 µS – 500 µS
Pulse rate:
100 Hz
.0 Hz to 999 Hz
Phase duration mod:
Modulation time:
Modulation:
OFF
ON or OFF
Auto Intensity:
10%
0 – 20%
Treatment time:
20 minutes
0 – 99 minutes
Channel B delay:
Ramp up:
Ramp down:
Time on:
Time off:
Channels A/B:
Waveform:
Asymetric Biphasic
Output:
0 mA
0 – 99 mA
ANALGESIA MOTOR:
Motor level stimulation activates motor and A-delta fibers causing the release of Bendorphin systemically. Duration
of relief is typically from 2-6 hours. Slow onset of relief usually within 30 minutes to 1 hour.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Phase duration:
250 µS
30 µS – 500 µS
Pulse rate:
4 Hz
.0 Hz to 999 Hz
Phase duration mod:
Modulation time:
Modulation:
OFF
ON or OFF
Auto Intensity:
10%
0 – 20%
Treatment time:
20 minutes
0 – 99 minutes
Channel B delay:
Ramp up:
Ramp down:
Time on:
Time off:
Channels A/B:
Waveform:
Asymetric Biphasic
Output:
0 mA
0 – 99 mA