OMNISTIM
®
500A USER MANUAL
29
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
IFC CUSTOM (FREQUENCY DIFFERENCE):
Protocol fully adjustable by therapist, which can be customized for deep tissue treatment.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
IFC Mode:
Frequency Difference
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
Burst rate scan:
Burst scan time:
Beat:
50 Hz
.0 – 250 Hz
Sweep:
100 Hz
.0 – 250 Hz
Rate Scan Time:
15 Seconds
0 – 20 Seconds
Rate scan:
Vector:
OFF
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
20%
0 – 20%
Treatment time:
20 minutes
0 – 99 Minutes
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA
5. Adjustable Parameters
Pushing the set button during operation brings up the available adjustable parameters. These may be altered during
system operation with the exception of the carrier frequency and mode of operation (Full Field or Frequency
Difference), which may only be altered when the output is turned off to the patient. (This prevents accidental shock
to the patient resulting from the change in pulse charge.) To adjust any parameter simply push the parameter set
button at any time during operation. This will display a parameter, which may be changed. Repeated pushing of the
parameter set button will continue to display parameters one at a time until all parameters have been displayed. To
change the parameter use the timer/parameter set rocker switch on the front panel to increase, decrease or change the
value of the parameter.
BEAT - SWEEP:
Beat adjustable from .1 - 250 Hz Sweep adjustable from beat to 250 Hz
RATE SWEEP TIME:
Adjustable from 1-20 seconds
CARRIER FREQUENCY:
Select 2000, 2500, 4000 or 5000 Hz.
VECTOR TECHNIQUE:
The vector scans the amplitude of each channel simultaneously. The modulation
may be set to sweep a 45 or 90-degree field. The sweep time of the vector can be set at fast (10 second up and
down sweep time) or Slow (60 second up and down sweep time).
AUTO INTENSITY:
Adjusts the output during treatment automatically from 0 to 20% user programmable.
IFC FD-Frequency Difference Interferential Mode:
This mode is generally used for treatment of deep tissue with nerve block effects occurring at the surface tissue.
Output channel A fixed at 2000, 2500, 4000 or 5000 Hz. Output channel B variable from 0.1 to 250 Hz Beat rate
and
sweep
may
be
set
during
treatment.