32 OMNISTIM
®
500A USER MANUAL
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
NMES - MUSCLE SPASM:
Activation of muscle with electrical stimulation with a 1:1 ratio of ON/OFF times causes muscle fatigue reducing
muscle spasm. Sensory stimulation of the muscles also reduces pain and thus reflex muscle spasm.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
80Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
8 seconds
0 – 20 Seconds
Rate scan:
ON
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
15 minutes
0 – 99 Minutes
Channel B delay:
Ramp up:
3 seconds
.0 – 9.9 seconds
Ramp down:
3 seconds
.0 – 9.9 seconds
Time on:
10 seconds
1 – 30 seconds
Time off:
10 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
ALTERNATE or
SIMULTANEOUS
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA
HIGH PERFORMANCE MFAC RE-ED STRENGTH:
Activation of muscle with electrical stimulation at a high intensity for a short time with long ON/OFF ramps reduces
atrophy and improves strength. The patient should participate to the extent possible by contracting during the
stimulation ON time.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
2.5 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz 5.0 KHz
Burst frequency:
75 Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0 – 50%
Burst scan time:
4 seconds
0 – 20 Seconds
Rate scan:
ON
ON or OFF
Auto Intensity:
0%
0 – 20%
Treatment time:
10 minutes
0 – 99 Minutes
Channel B delay:
.0 seconds
0 – 99 Minutes
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
10 seconds
1 – 30 seconds
Time off:
50 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
B DELAYED, ALTERNATE
or SIMULTANEOUS
Waveform:
Alternating Current
Output:
0 mA
0 – 99 mA