OMNISTIM
®
500A USER MANUAL
45
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
LVPC (LOW VOLTAGE PULSED CURRENT) MODE:
Waveform:
Biphasic square wave
Phase Duration:
10-500µs
Pulse Rate:
0.1 - 999pps
HVPC (HIGH VOLTAGE PULSED CURRENT) MODE:
Waveform:
Twin monophasic pulses each with 40µs phase duration, with inter phase interval of 100µs
Pulse Rise Time:
<500 ns
Pulse Rate:
Adjustable from 1-125pps
Interphase Interval:
Adjustable from 10-250µs
TIMER FUNCTIONS:
Treatment Timer:
Adjustable for 1-99 minutes in one minute increments. Turns output to zero and system off and sounds
10-second buzzer to indicate completion of treatment.
Channel Timing:
In alternating mode channels follow each other sequentially. In simultaneous modes the output is
simultaneous. In delayed mode channel B is delayed.
TIMING SELECTIONS:
Ramp Time:
Adjustable from 0.1 to 9 seconds in 0.1-second increments. Ramps overlap in NMES Alt mode or with
channel delay
ON Time:
Adjustable from 1 to 30 seconds in 1-second increments
OFF Time:
Adjustable from 0 (continuous) to 199 seconds in 1-second increments
PATIENT SAFETY SYSTEMS:
Activation:
Patient safety hand control shuts down output. Output modality may not be changed during operation. Output
levels are reset to zero at the start and completion of treatment.
MISC.:
Auto Intensity:
Output intensity is gradually increased during treatment from 0 to 20% automatically in a linear manner
over the duration of the treatment when the anti-adaptation mode is selected. The amount of the increase
is user programmable.
Treatment Programs:
* For program details see page 24.
IFC:
5 Presets
1 Custom
MFAC:
6 Presets
1 Custom
LVPC:
7 Presets
1 Custom
HVPC: 2
Presets
Certificates and Approvals:
Devices are designed to meet or exceed all safety requirements of a medical device in its class per IEC
60601 and CSA C22.2 No. 601.1
Caution:
Federal law restricts this device to sale by or on the order of a physician (or other health practitioner
licensed by their State).
ACP reserves the right to change technical specifications and product availability without notice.