LVPC CUSTOM:
LVPC protocol which can be customized by the therapist.
PARAMETER DEFAULT
VALUES
RANGE OF USER
ADJUSTABLE VALUES
Phase duration:
200 µS
30 µS – 500 µS
Pulse rate:
100 Hz
.0 Hz to 999 Hz
Phase duration mod:
0 %
0 – 50%
Modulation time:
0 seconds
0 – 20 seconds
Modulation:
OFF
ON or OFF
Auto Intensity:
20 %
0 – 20%
Treatment time:
20 minutes
0 – 99 minutes
Channel B delay:
0.0 seconds
.0 – 9.9 seconds
Ramp up:
2 seconds
.0 – 9.9 seconds
Ramp down:
2 seconds
.0 – 9.9 seconds
Time on:
5 seconds
1 – 30 seconds
Time off:
30 seconds
0 – 199 seconds
Channels A/B:
SIMULTANEOUS
B DELAYED, ALTERNATE
or SIMULTANEOUS
Waveform:
Asymetric Biphasic
Output:
0 mA
0 – 99 mA
2. Adjustable Parameters
Pushing the set button during operation brings up the available adjustable parameters. These may be altered
during system operation. To adjust any parameter simply push the parameter set button at any time during
operation. This will display a parameter which may be changed. Repeated pushing of the parameter set
button will continue to display parameters one at a time until all parameters have been displayed. To
change the parameter use the timer/parameter set rocker switch on the front panel to increase, decrease or
change the value of the parameter.
PULSE RATE:
Adjustable from .1 - 999 Hz
PHASE DURATION:
Adjustable from 10 to 500 microseconds.
PHASE DURATION MODULATION:
+/- 20% for a five second sweep time. Select on or off.
ON-OFF
RAMPS:
Independently adjustable from 0 to 9.9 seconds
ON - OFF TIME:
On time adjustable from 0 - 30 seconds, Off time adjustable from 0 -199 seconds.
CHANNEL TIMING:
Simultaneous, alternate or B channel delayed.
DELAY MODE:
Adjustable from 0-9.9 seconds
AUTO INTENSITY:
Adjusts the output during treatment automatically from 0 to20% user programmable.
LOW VOLTAGE PULSED CURRENT WAVEFORM
LVPC is generally used for surface and smaller muscle
stimulation and pain control with TENS technique.