Operation
Page 57
There are eight different pins available, which will be set high during an associated event:
DO N (Pin X)
Associates a state with a digital output. There are eight digital
outputs (the pin numbers are for the 25 pin connector). Each
one can have one of the associated states listed in Table 3 –
Digital Output States. The pin will be driven to the active state
while the instrument state is true.
Active
[DO Enabled]
Each pin can be set to be active
High
or
Low
. Active High
means that the pin will be pulled to 5V when the associated
event occurs. Active Low means the pin will be pulled to 0V
when the associated event occurs.
Table 3 – Digital Output States
Digital Output State
Description
Disabled
No state (this state is never Active).
Pwr Supply Fail
Power supply fault.
A2D Fail
Analog to digital fault.
Flow Fail
Sample flow fault.
Cell Heat Fail
Cell heater fault.
Flow Heat Fail
Flow block heater fault
Chassis Tmp Fail
Chassis temperature fault.
USB Disconnected
The USB memory stick is disconnected.
Background
Performing a background.
Span
Performing a span check.
Zero
Performing a zero check.
System Fault
Any system fault (the red light is on).
Maintenance Mode
User has activated maintenance mode.
Over Range AO 1
Over range for analog output is active.
Interface PCA Fail
Interface PCA fault.
High Background
Background above 1100 Mm-1
Signal Fail
Signal fault.
3.4.28
Network Menu
Main Menu
Communications Menu
Network Menu
The
Network Menu
only appears when the
Network Port
is enabled in the Hardware Menu (refer to
Section 3.4.31). The
Network Menu
allows the user to view or set the IP Address, Netmask and
Gateway if the optional network port is installed.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...