Appendix B.
EC9800 Protocol
The Serinus implements a subset of the 9800 instrument protocol. Only the basic commands of
reading the concentration value and setting the instrument calibration state (measure, span or zero)
are supported.
B.1
Command Format
All commands are sent as ASCII strings. Fields are delimited by commas and the command ends with
the normal return key (i.e. the TERMINATOR is either a <CR> or a <LF>). The DEVICE I.D. is the Serial
ID assigned in the
Main Menu
Communications Menu
Serial Communication Menu
. If the
instrument is not being used in a multi-drop connection, the DEVICE I.D> can be replaced with the
string “???”.
B.2
Commands
B.2.1
DCONC
Function:
Sends the current instantaneous concentration data to the serial port.
Format:
DCONC, {<DEVICE I.D.>} {TERMINATOR}
Device response:
{GAS} <SPACE> {STATUS WORD} <CR><LF>
The GAS value is the concentration value in the current instrument units, expressed as a floating
point number (i.e. 12.345). The STATUS WORD indicates the instrument status in hex (i.e. A01F) using
the following format:
Bit 15 = SYSFAIL (MSB)
Bit 14 = FLOWFAIL
Bit 13 = LAMPFAIL
Bit 12 = CHOPFAIL
Bit 11 = CVFAIL
Bit 10 = COOLERFAIL
Bit 9
= HEATERFAIL
Bit 8
= REFFAIL
Bit 7
= PS-FAIL
Bit 6
= HV-FAIL
Bit 5
= OUT OF SERVICE
Bit 4
= Instrument is in zero mode
Bit 3
= Instrument is in span mode
Bit 2
= Unused
Bit 1
= SET
PPM selected, CLEAR
MG/M3
Bit 0
= reserved (LSB)
B.2.2
DSPAN
Function:
Commands the instrument to enter span mode.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...