Serinus 60 User Manual 3.0
Page 96
recommends the use of a dilution calibrator and a certified cylinder of NO
2
at an appropriate
concentration (typically a cylinder of 40 to 100 ppm, balance in Nitrogen).
Equipment Required
Gas Dilution Calibrator (such as a Serinus Cal 1000).
A NO
2
gas cylinder reference standard
Zero Air Generator
Procedure
1.
Connect your calibration system to the
Calibration Port
of the instrument (Ecotech recommends
the Serinus Cal 1000, refer to Figure 43).
2.
Generate and record the displayed span concentrations for (at least) five stepped points (of
known concentrations) evenly spaced across the instruments measurement range (see the
example below).
3.
Then using a program such as MS Excel, create an X Y scatter plot of expected concentration
versus the recorded instrument response and use linear regression to calculate the line of best fit
and the correlation factor (R
2
) – refer to the users local applicable standards.
Example for an instrument measurement range of 500 ppb:
a.
For the 1st concentration, set the gas dilution calibrator to supply
400 ppb
NO
2
gas to the
instrument.
b.
Allow the instrument to sample the calibration gas until a prolonged stable response is
achieved (the amount of time this takes is impacted by the calibration setup). Record the
instrument response.
c.
Repeat the above steps using concentrations of
300ppb
,
200ppb
,
100ppb
and a
Zero
point.
d.
Graph the results and use linear regression to determine a pass or fail as per your applicable
local standards.
𝑦 = 𝑚𝑥 + 𝑐
Note:
To highlight hysteresis errors, it is advisable to run the multipoint check in both
descending
and
ascending
order - refer to applicable local standards.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...