Serinus 60 User Manual 3.0
Page 30
Items Received
With the delivery of the Serinus 60, the user should have received the following:
Ecotech Serinus 60 Analyser
PN: E020060
Green Ecotech Resources USB Stick
PN: H030137-01
Manual
PN: M010060 (hardcopy optional)
USB Memory Stick
PN: H030021
USB Cable
PN: COM-1440
Power Lead (120V)*
USA
PN: C040007
Power Lead (240V)*
Australia
PN: C040009
Europe
PN: C040008
UK
PN: C040010
*The power lead received depends on the mains supply of the country (120V or 240V).
Note:
Check all these items have been delivered undamaged. If any item appears damaged, please
contact your supplier before turning the instrument on.
2.2
Installation Notes
When installing the instrument the following points must be taken into account:
The instrument should be placed in an environment with minimal dust, moisture and variation in
temperature (20-30 °C for U.S. EPA designated range).
For best results the instrument should be located in a temperature and humidity controlled
environment (air conditioned shelter). An enclosure temperature of 25-27 °C is optimum.
Whether in a rack or placed on a bench, the instrument should not have anything placed on top
of it or touching the case.
Instruments should be sited with easy access to the front panel (instrument screen/USB memory
stick) and to the back panel (communication ports/pneumatic connections).
It is recommended that the sample line be as short as possible and/or a heated manifold be used
for sampling (minimising moisture condensation in the sample).
Do not pressurize the sample line under any circumstances. Sample should be drawn through the
instrument under atmospheric pressure. This should be achieved by using an external vacuum
pump connected to the exhaust port of the instrument.
When supplying span gas, ensure the flow is approximately 1 slpm and excess is sufficiently
vented.
Note:
The power on/off switch is accessible from the rear of the instrument only. Install the
instrument so that the on/off power switch is accessible.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...