Serinus 60 User Manual 3.0
Page 44
Set Point (FLOW)
[Internal Pump Option]
The temperature set point of the flow block. The factory default is
50 °C.
Cell
Displays current temperature of the measurement cell.
Flow Block
[Internal Pump Option]
Displays current temperature of the flow block.
Chassis
Displays the temperature of air inside the chassis, measured on
the main controller PCA.
3.4.6
Pressure & Flow Menu
Main Menu
Analyser State Menu
Pressure & Flow Menu
Pressure Units
Select the units that the pressure will be displayed in (
torr
,
PSI
,
mBar
,
ATM
or
kPa
).
Ambient
Current ambient pressure.
Cell
Current pressure within the measurement cell.
Internal Pump
[Internal Pump Option]
Current pressure within the flow block.
Flow Units
Select the units that the sample flow will be displayed in (
slpm
or
L/min
).
Sample Flow
Indicates the gas flow through the sample port of the instrument.
The value should be ~0.50 slpm. If there is an error with the
sample flow, it will read 0.00 slpm.
3.4.7
Voltage Menu
Main Menu
Analyser State Menu
Voltage Menu
Signal Voltage
This is the signal voltage reading.
+5V Supply
+5 V power supply.
+12V Supply
+12 V power supply.
+Analog Supply
+12 V (primary) power supply. The value should be within ±2 V.
-Analog Supply
-12 V (primary) power supply. The value should be within ±2 V.
3.4.8
General Settings Menu
Main Menu
General Settings Menu
Decimal Places
Select the number of decimal places (0-5) used for the
concentration displayed on the home screen.
Conc. Units
Sets the concentration units (
ppm
,
ppb
,
ppt
,
mg/m3
,
μg/m3
or
ng/m3
).
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...