Serinus 60 User Manual 3.0
Page 90
9.
Turn off the vacuum source and disconnect the barometer from the
Sample Port
. Wait ~1
minutes for the pressure to equalise within the instrument.
10.
Disconnect the vacuum source from the
Exhaust Port
.
11.
Wait 2-5 minutes, enter the ambient barometer reading into the instrument - Accept.
Note:
Both of the pressure sensors should now be displaying the current ambient pressure and they
should be the same value within 3 torr of each other.
12.
Back -
Pressure Calibration Menu
- (read note) - OK.
5.2.2
Ambient Point Pressure Calibration
Full pressure calibrations are generally recommended, however it is possible to calibrate only the
ambient point in cases where only a minor ambient pressure adjustment is required.
Note:
Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure the instrument’s stability.
Note:
Ensure units of measure are the same on both the barometer and instrument.
Equipment Required
Barometer
Procedure
1.
Open -
Main Menu
Calibration Menu
Pressure Calibration Menu
- (read note) - OK.
Note:
This action will place the valve sequence on hold; normal sampling will be interrupted.
2.
Edit -
Ambient Set Pt.
- (Read displayed instructions) - OK.
3.
Disconnect any external tubing connected to the rear ports of the instrument (
Sample Port
,
Calibration Port
and
Exhaust Port
).
4.
Wait 2-5 minutes and enter the ambient barometer reading into the instrument - Accept.
Figure 44 – Setting the Ambient Set Point
Note:
Both of the pressure sensors should now be displaying the current ambient pressure and they
should be the same value within 3 torr of each other.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...