A.2.2
Get IEEE Value
Where:
Command byte
1
Message byte 1
Index from List of Parameters
Message byte 2..32 Additional indexes (optional)
This command requests the value of an instrument parameter. The message field byte contains the
index of the parameter requested, as described in the List of Parameters.
Up to 32 indexes can be supplied in a single request. The response has 5 bytes for each parameter
requested – the first byte is the parameter index and the next four are the IEEE representation of the
current value.
Example
A request with a message field of 50,51,52 to a Serinus S40 would return a 15 byte message as
shown below:
Table 18 – Example: Get IEEE Response data
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
50
NO reading
51
NOx reading
52
NO2 reading
A.2.3
Set Calibration Mode
Where:
Command byte
4
Message byte 1
85
Message byte 2-5
The IEEE representation of 0, 1, 2, or 3
0 puts the instrument into Measure mode (0,0,0,0)
1 puts the instrument into Cycle mode (63,128,0,0)
2 puts the instrument into Zero mode (64,0,0,0)
3 puts the instrument into Span mode (64,64,0,0)
This command puts the instrument into a calibration mode (the same as going to the Calibration
menu and choosing a Cal. Mode).
Example
A request with a command of 4 and a message field of 85,64,64,0,0 would place the instrument into
Span mode.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...