6.4.3
Updating Firmware
It is important for optimal performance of the instrument that the latest firmware is loaded. The
latest firmware can be obtained by visiting Ecotech’s website:
http://www.ecotech.com/downloads/firmware
Or by emailing Ecotech at
To update the firmware from a USB memory stick, use the following procedure:
USB Memory Stick Update
1.
Turn the instrument off.
2.
Place the USB memory stick with the new firmware (ensure that firmware is placed in a folder
called FIRMWARE) in the front panel USB Port.
3.
Enter the Bootloader (refer to Section 6.4).
4.
Select option 3 (upgrade from USB memory stick), press 3 on the keypad.
5.
Wait until the upgrade has completed.
6.
Press 9 on the keypad to start the instrument with new firmware.
6.4.4
Erase All Settings
This command is only required if the instrument’s firmware has become unstable due to corrupted
settings. To execute this command enter the
Bootloader Menu
(refer to Section 6.4) and press 4 on
the keypad.
6.4.5
Start Analyser
The start analyser command will simply initiate a firmware load by pressing 9 on the keypad from the
Bootloader Menu
. It is generally used after a firmware upgrade.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...