Operation
Page 59
Bluetooth
This field indicates whether the instrument is remotely
connected to an Android device.
Reset
After changing the ID or PIN, it is necessary to reboot the
Bluetooth module. This is done by resetting the instrument or
by using this menu item to reboot only the Bluetooth.
ID
This is the Bluetooth ID of the instrument. Use the keypad to
edit this field (refer to Section 3.3.1 for instructions on
entering text with the numeric keypad). The default ID setting
is
Serinus
(
Ecotech ID
). The word
Serinus
is always the first
part of the name and cannot be edited. The second part is the
Ecotech ID
.
PIN
This is a passcode/pin required for the Serinus Remote
Application to connect to the instrument. The default pin is
1234.
3.4.30
Advanced Menu
This menu is accessed via a different method than the other menus. From the
Home Screen
press
the following keys:
(
.
̅
)99(
+
SPACE
)
This menu contains technical settings, diagnostics and factory hardware installations. No items in this
menu should be accessed without authorisation and supervision of qualified service personnel.
Language
Select a language.
Hardware Menu
Refer to Section 3.4.31.
Service Displays
When set to
On
, new items appear on many different menus.
These fields are for diagnostic and service personnel only.
Default is
Off
.
Next Service Due
Enables the user to edit the next service due date.
Jump to Next State
Moves the sequence to the next state (e.g. from Fill to
Measure). This command is most commonly used to force an
instrument out of the warm-up sequence early.
Parameter Display Menu
Refer to Section 3.4.32.
Reset to Factory Defaults
Reset the configuration to factory defaults. This will erase all
calibrations and user configuration information.
Rebuild Index
If a data log becomes corrupted it may be possible to restore
It by rebuilding its index file. This command will ask you to
specify a month and will rebuild the index for that month. This
operation can take many minutes and it should not be
interrupted. While the file is rebuilding any data logging will
be suspended.
3.4.31
Hardware Menu
Advanced Menu
Hardware Menu
This menu contains factory hardware installations. If you reset to factory defaults you may need to
revisit this menu to enable your installed optional features.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...