8 . . .Diagnosis / error messages
. . .Error messages
Error no . / Range
Text on the LCD display
Cause
Remedy
F086.019 / Electronics
Curr.Out 31 / 32 com error.
Defective Board. EMC
disturbance. Call Service.
Defective Motherboard Hardware. EMC
disturbance.
Call Service.
F084.011 / Electronics
NV data defect. Data storage
irreparable. Call Service.
Data in SensorMemory corrupt.
Call Service.
F082.014 / Electronics
Incompatible Frontend Board.
Frontend not fit to Motherboard.
Call Service.
Wrong Frontend or Mother Board.
Call Service.
F081.026 / Electronics
MB voltages outside range.
Defective Motherboard HW. Call
Service.
Defective Mother Board Hardware.
Replace Mother Board.
Call Service.
Error no . / Range
Text on the LCD display
Cause
Remedy
C078.004 / Config.
Flowrate to zero.
Check digital in terminals.
The option card Digital Input is configured to
"set flowrate to zero" and this event is
triggered.
Check Digital Input terminals and
configuration.
C076.006 / Config.
All totalizer stopp.
Check digital in terminals.
The option card Digital Input is configured to
"All Totalizer stopp" and this event is
triggered.
Check Digital Input terminals and
configuration.
C074.007 / Config.
Totalizer reset.
Reset of one or more Totalizers.
The option card Digital Input is configured
to "ResetTotalizer" and this event is
triggered.
Check Digital Input terminals and
configuration.
C072.003 / Config.
Simulation is on Simulated
values.
Switch off Simulation Mode.
The simulation of a process value or an
output is active.
Switch off simulation mode.
C070.027 / Config.
An alarm is simulated.
Switch off alarm simulation.
The simulation of an alarm is active.
Switch alarm simulation to "Off".
Function check