Menu / parameter
Device Setup / . . .Sensor / . . .Operating Mode
Flow Direction
Sets the measuring direction for the sensor.
As delivered, the device measures and counts in both flow directions.
• Forward & Reverse: The device measures in both flow directions.
• Forward only: The device measures only in the forward flow direction.
• Reverse only: The device measures only in the reverse flow direction.
Default setting: Forward & Reverse
Flow Indication
Inverts the flow direction displayed.
Default setting: Normal
7 . . .Operation
. . .Parameter descriptions
Device Setup / . . .Transmitter
. . .Units
Selection of submenu ‘...
Units
’ using .
Damping
Select the damping.
The value (0.02 ... 60 s) set here relates to 1 Τ (Tau). The value refers to the response time for a step flowrate
change. It affects the instantaneous value in the display and at the current output.
Default setting: 1 second
Damping On/Off
Switches the damping on or off.
Density Fixed Value
If the flow is measured and indicated in mass flow units, a fixed density must be taken into account for the
calculations. To convert the flow rate to mass flow units, a density value from 0.01 ... 5.0 g/cm3 can be entered.
. . .Cut Off
Selection of submenu ‘...
Cut Off
’ using .
TX Location TAG
Entry of the measuring point tag for the transmitter.
Alphanumeric, max. 20 characters
TX TAG
Enter the tag number for the transmitter.
Alphanumeric, max. 20 characters
Plant Data Sync.
Tx -> Sens
The settings are redundantly saved in two data modules. One of them is the SensorMemory, the other is the
transmitter motherboard (backplane).
By selecting ‘Tx -> Sens’, location-specific settings such as measuring range or damping are replicated from
the transmitter motherboard (backplane) to the SensorMemory.
Sens -> Tx
By selecting ‘Sens -> Tx’, location-specific settings such as measuring range or damping are replicated from
the SensorMemory to the transmitter motherboard (backplane).
Device Reset
For service purposes only. Restarts the device without the need for cycling the power.
Restore Factory Def.
All user-accessible parameters will be reset to the factory default settings.
. . .Feature Settings
Selection of submenu ‘...
Feature Settings
’ using .
. . .System Zero
Selection of submenu ‘...
System Zero
’ using .
Noise Reduction
Activates the filter technology for noise reduction.
Filter: Off, Filter 15, 30, 60 (15: lower filtering, 60: strong filtering)
Filter setting affects 20 mA signal (damping).
Default setting: Off
Menu: Device setup