Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
7 CALIBRATION
IM/AX4CO Issue 11
63
Factory Settings Access Code
Enter the required code number (between 0000 and 19999) to gain access to the
factory settings. If an incorrect value is entered, access to subsequent frames is
prevented and the display reverts to the top of the page.
Calibrate Sensor A
Note.
The values in the display lines for sensor calibration are shown only as examples
- the actual values obtained will differ.
Sensor B calibration (dual input analyzers only) is identical to Sensor A calibration.
Single input analyzers only - see page 66.
Operating Page
Calibrate Input for Sensor A ?
If calibration is required select
Yes
otherwise select
No
.
Note.
To abort calibration, press the
key again at any time before calibration is
complete - see next page.
Continued on next page.
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A: Res.Zero (O/C)
Cal. Sensor B
Cal. Output 1
Conductivity
Calibrate I/P ?
-----
Yes
No
Yes
No
Dual Cond.