Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
5 PROGRAMMING
24
IM/AX4CO Issue 11
Low Range Temperature Compensation
Select the type of low range (0 to 100ºC [32 to 212ºF]) temperature compensation
required:
Linear
UPW*
Acid**
NaOH***
NaCl*
NH3**
-
-
-
-
-
-
Linear temperature compensation based on a manually entered
temperature coefficient (see appendix A.1 on page 73) - see
Temp.Coeff.
frame on page 26.
Example
- Non-standard applications.
Temperature compensation based on the temperature coefficient
of pure water. Source data is based on International Standard IEC
60746-3.
Also enables manual entry of a temperature coefficient (see
Temp.Coeff.
frame on page 26) for applications where pure water
contains an unknown impurity; in which case the temperature
coefficient must be calculated - see appendix A.1.1 on page 74.
Temperature compensation based on the temperature coefficient
of pure water containing trace acids.
Examples
- Cation exchanger in-bed and outlet applications.
- Degassed cation/conductivity applications.
Temperature compensation based on the temperature coefficient
of pure water containing trace caustic.
Example
- Inferred pH in caustic-dosed waters applications.
Temperature compensation based on the temperature coefficient
of pure water containing trace salts.
Examples
- General monitoring applications.
- Mixed-bed exchanger applications.
- Final polisher effluent applications.
- Cation exchanger inlet applications.
- Anion exchanger in-bed and outlet applications.
- Reverse osmosis applications.
Temperature compensation based on the temperature coefficient
of pure water containing trace ammonia.
Examples
- Ammonia-treated make-up and boiler feed water applications.
- Condenser sampling applications.
- Hot well sampling applications.
- Before-cation column applications.
- Inferred pH in ammonia-dosed waters applications.
*
**
***
Applicable only on conductivities up to 10
μ
S cm
-1
Applicable only on conductivities up to 25
μ
S cm
-1
Applicable only on conductivities up to 100
μ
S cm
-1
Continued on page 26.
A: Temp.Sensor
A: Temp.Comp.
-----
NH3
NaCl
NaOH
Acid
UPW
Linear
A: T.Comp Range
set to
Lo TC