7 CALIBRATION
66
IM/AX4CO Issue 11
Calibrate Output 1
Note.
When adjusting the 4 and 20mA output values, the display reading is
unimportant and is used only to indicate that the output is changing when the
and
keys are pressed.
See below.
Adjust 4mA
Use the
and
keys to set the milliammeter reading to 4mA.
Note.
The analog output range selected in
Configure Outputs
(Section 5.5) does not
affect the reading.
Adjust 20mA
Use the
and
keys to set the milliammeter reading to 20mA.
Note.
The analog output range selected in
Configure Outputs
(Section 5.5) does not
affect the reading.
See below.
Operating Page
Calibrate Output 2
Note.
Output 2 calibration is identical to Output 1 calibration.