5 PROGRAMMING
48
IM/AX4CO Issue 11
5.11 Test Outputs and Maintenance
Test Outputs
Displays the output test details for the analog outputs.
Note.
Outputs 3 and 4 are available only if the option board is fitted
and
analog features
enabled - see section 7.3, page 62.
Test Output 1
frame only is shown; the format of frames for the remaining outputs is
identical.
See below.
Test Output 1
The theoretical output current value.
Output current as a percentage of the full range current.
Use the
and
keys to adjust the displayed theoretical output current value
to give the output required.
Test remaining outputs.
Maintenance
Hold Outputs
Enables the relay action and analog outputs to be maintained.
Auto.
On
Off
-
-
-
Changes in relay action and analog outputs are disabled during
sensor calibration.
Changes in relay action and analog outputs are disabled.
Changes in relay action and analog outputs are enabled.
Note.
The LEDs flash while the analyzer is in 'Hold' mode.
Continued on next page.
Hold Outputs
set to
Off
or
On
- return to main menu.
Hold Outputs
set to
Auto.
- continued on next page.
Test Outputs
-----
TEST/MAINTENANCE
-----
Test Output 1
4.00
mA
20.0
%
Maintenance
-----
Hold Outputs
-----
Auto.
On
Off
FACTORY SETTINGS
Test Output 2
FACTORY SETTINGS
Load/Save Config
Maintenance
Maintenance
Automatic Time