Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
5 PROGRAMMING
26
IM/AX4CO Issue 11
Temperature Sensor
Select the type of temperature sensor used, Pt100 or Pt1000.
Temperature Coefficient
Notes.
•
•
Displayed only if
T.Comp Range
is set to
Lo TC
and
Temp.Comp.
is set to
Linear
or
UPW
or
T.Comp Range
is set to
Hi TC
and
Temp.Comp.
is set to
UPW
- see pages
23 to 25.
If
A: Cond Units
is set to
USP645
(AX450 and AX455 analyzers only - see page 22),
the temperature coefficient is fixed automatically at 2.00%/ºC.
Enter the temperature coefficient (a x 100) of the solution (0.01 to 5.0%/ºC). If unknown,
the temperature coefficient (a) of the solution must be calculated - see appendix A.1.1
on page 74.
If the value has not yet been calculated, set it to 2%/ºC provisionally.
TDS Factor
Note.
Displayed only if
A: Cond.Units
is set to TDS - see page 22.
The TDS factor must be programmed to suit the particular application - see appendix
A.2 on page 74.
Enter the required TDS factor between 0.4 and 0.8.
For salinity applications, set the TDS factor to 0.6.
TDS Units
Note.
Displayed only if
A: Cond.Units
is set to
TDS
Select the TDS units (ppm, mg/l or mg/kg).
Enable Calibration
Select
Yes
to enable sensor calibration - see section 4.1, page 17.
If
No
is selected the calibration menu, pages and frames for the relevant sensor are not
displayed.
Sensor B configuration (dual input analyzers only) is identical to Sensor A configuration.
Single input analyzers only - return to main menu.
A: Temp.Sensor
-----
A: Temp.Coeff.
2.00
Pt100
Pt1000
%/
°
C
A: TDS Factor
0.75
ppm/uS
-----
A: TDS Units
mg/kg
mg/l
ppm
-----
A: Enable Cals.
Yes
No
Config. Sensor B
CONFIG. ALARMS
Config. Sensor A
CONFIG. SENSORS