34
5 Confirm that there is nothing on the pan and press
the PRINT key. The balance measures the zero
point. Do not allow vibration or drafts to affect the
balance.
The balance displays the calibration weight value.
6 Place the displayed calibration weight on the pan
and press the PRINT key. The balance measures
the calibration weight. Do not allow vibration or
drafts to affect the balance.
7 The balance displays
end
. Remove the weight from
the pan.
8 When the GLP output is set, the balance displays
1
glp
and outputs “Calibration Report”.
9 The balance will automatically return to the weighing
mode.
10 Place the calibration weight on the pan and confirm
that the value displayed is within
±
2 digits of the
specified value. If it is not within the range, check the
ambient conditions such as drafts and vibration, also
check the weighing pan. Then, repeat steps 1 to 9.
Summary of Contents for FZ-i WP Series