43
Class
Item and Parameter
Description
Cp Hi
Upper limit
Refer to “10.10. Comparator
Function”.
Cp lo
Lower limit
dout
Data output
prt
Data output mode
0
Key mode
Accepts the PRINT key only
when the display is stable.
1
Auto print mode A
(Reference = zero)
Outputs data when the display
is stable and conditions of
ap-p
,
ap-b
and the reference
value are met.
2
Auto print mode B
(Reference = last stable value)
3
Stream mode
Outputs data continuously.
4
Key mode B (Immediately)
Accepts the PRINT key regardless of
the display condition.
5
Key mode C (When stable)
Accepts the PRINT key immediately
when the display is stable, or waits for
the display to be stable when not.
6
Interval output mode
Uses interval output mode.
ap-p
Auto print polarity
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
2
Both
Regardless of displayed value
ap-b
Auto print difference
0
10 digits
Difference between reference
value and displayed value
1
100 digits
2
1000 digits
int
Interval time
0
Every measurement
Interval time for the interval
output mode
(With
prt 6
)
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minute
7
5 minute
8
10 minute
5-id
ID number output
0
No output
Selects whether or not the ID
number is output.
1
Output
5-td
*
Time/Date output
0
No output
1
Time only
2
Date only
3
Time and date
pU5e
Data output pause
0
No pause
Selects the data output
interval.
1
Pause (1.6 seconds)
at-f
Auto feed
0
Not used
Selects whether or not
automatic feed is performed.
1
Used
info
GLP output
0
No output
Selects GLP output method.
1
AD-8121 format
2
General data format (output time and date using
the built-in clock)
3
General data format (output time and date using
the clock of the external equipment)
ar-d
Zero after output
0
Not used
Adjusts zero automatically
after data is output
1
Used
Factory settings. Digit is a unit of minimum weighing value
*