77
T command example
PT command example
When a communication error has occurred due to external noise, or a parity error has occurred
due to transmission error, the balance transmits an error code. In this case, send the command
again.
18.3. Settings Related to RS-232C
Concerning the RS-232C, the balance has two functions: “Data output (
dout
)” and “Serial interface
(
5if
)”. Set each function as necessary.
PT command
(Tare of 100g)
?PT command
LF
CR
T
P
?
PT:+ 00100.00 g
CR LF
P T
:
1 0 0
Tare value display
AKCR LF
g
CR LF
AKCR LF
Personal computer Time Balance
Weighing mode
Command completion
?PT command
T
CR LF
AKCR LF
AKCR LF
LF
CR
T
P
?
PT:+ 00126.87 g
CR LF
T command
AKCR LF
AKCR LF
Personal computer Time Balance
Weighing mode
Zero display
Processing command
Confirmation of
command receipt
Command completion