Immunity test
Compliance to IEC 60601 test levels Electromagnetic environment – guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
see following table:
Test specifications for enclosure port
immunity to RF wireless communica-
tions equipment
Electrical fast tran-
sients / bursts
IEC 61000-4-4
± 2 kV for power supply lines
100 kHz repetition frequency
Mains power quality should be that of a
typical commercial or hospital environment.
Conducted distur-
bances induced by RF
fields
IEC 61000-4-6
3 V
rms
0,15 MHz - 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Surges
Line-to-line
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, and
short interruptions on
power supply input
lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45 °, 90°, 135°, 180°, 225°,
270°, 315°
0 % U
T
; 1 cycle and
70 % U
T
; 25130 cycles
Single phase: at 0°
0 % U
T
; 2501300 cycle
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Elipar DeepCure-L Dental
Light Unit requires continued operation during
power mains interruptions, it is recommended
that the Elipar DeepCure-L Dental Light Unit
be powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE
U
T
is the a. c. mains voltage prior to application of the test level..