Page 54
User manual 5448X000 Version 1 June-06
Printing out
During measurement, users can print out the memorized measured results at any time
when
(F4)
key is pressed.
Quitting measuring operation
After measurement, press
Return
key
(4.3e). key to quit measuring operation.
The system indicates that the data will
be lost. Users can press
Return
key (4.3e)
key to return to the main menu of pho-
tometric measure.
Spectrum Scanning
This mode provides a wavelength scan and print function for absorbance, transmittan-
ce and energy of the measured sample.
Information display and function keys
On the instrument main menu (mode menu), press
(2)
key to select spectrum scanning
mode. The main menu of spectrum scanning mode is displayed on LCD, as shown in pic-
ture 18.
The first line on the menu indicates
that the current working mode is spec-
trum scanning. In the upper part of ope-
rating area displays the current working
wavelength (660.0nm) and the measu-
red value (0.000Abs), while in the lower
part is a coordinate diagram.
On the right, the function of each
function key is displayed.
(F1):
Parameter setting
(F2):
Peak-picking
(F3):
Sample control
(F4):
Spectrum printing
On this page, wavelength setting, baseline correction and spectrum scanning can be
done by pressing
GoTo WL
key (4.4b),
Autozero
key (4.4a) and
Start/Stop
key (4.4c) res-
pectively.
14:57
Print OK?
Yes: Enter , No: CLS
PARM
(F1)
CLS
(F2)
SAMP
(F3)
PRIN
(F4)
Photo
660.0
0.000
No.
Abs
K*Abs
1.
0.000 0.000
2.
0.000 0.000
3.
0.001 0.001
Picture 17
14:57
System Ready
PARM
(F1)
PEAK
(F2)
PRIN
(F4)
Spectrum
Abs
200
250
300
350
400
-0.010
-0.005
0.000
0.005
0.010
660.0nm
0.000
Picture 18
Version 1 June-06
User manual 5448X000 Page 59
Information display and function keys
On the instrument main menu, press
(3)
key to select the quantitative measu-
rement mode. The main menu of quan-
titative measurement will be displayed
as shown in picture 27.
The first line on the menu is the mode
line, which shows that the current opera-
tion mode is quantitative measurement.
The upper part of the operating area
shows the current measuring wave-
length (500.0nm) and the measured
value (0.000Abs). The lower part is a
chart in which measured data is displayed.
The right part of the page is the function key indicating area, which shows function keys
(F1) ~(F4)
.
(F1):
Parameters setting
(F2):
Clear measured data
(F3):
Sample control
(F4):
Printing
On this page the measuring wavelength setting, autozero-ing and sample measurement
can be carried out by pressing
GoTo WL
key (4.4b),
Autozero
key (4.4a) and
Start/Stop
key
(4.4c) respectively.
Parameter setting
Parameter setting page can be displa-
yed by pressing
(F1)
key on the main
menu of quantitative measurement, as
shown in picture 28.
On this page the measurement
method, measuring wavelength and
concentration unit can be set directly by
pressing the numeric keys
(1) ~ (3)
. By
pressing twice
(F1)
key the calibration
curve setting page can be displayed.
After finishing all parameter settings, press
Return
key (4.3e) to return to the previous
page.
Setting measurement method
Press
(1)
key to set measurement method for quantitative measurement. There are two
selections: K-factor method and calibration curve method. The two methods can be selec-
ted by pressing
(1)
key alternatively.
14:57
System Ready
PARM
(F1)
CLS
(F2)
SAMP
(F3)
PRIN
(F4)
Quant.
660.0
0.000
No.
Abs
Conc.
Picture 27
14:57
Quantitation:
Parameter
1. Measure method: Standard
2. Measure WL:
660.0
3. Conc. Unit:
mg/l.
Select Item No.:
PARM
(F1)
Picture 28
ENGLISH
ENGLISH