![Zuzi UV-VIS 4481 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/zuzi/uv-vis-4481/uv-vis-4481_manual_942780039.webp)
Page 74 User manual 5448X000 Version 1 June-06
Open the light source housing cover. Loosen the two holding screws (53.1) on the W
lamp socket and remove the halogen lamp holder (W) .
Replace the halogen lamp with a new one. Clean the surface of the W lamp with a clean
cloth dampened with alcohol.
Insert the halogen lamp into the halogen lamp socket in the light source housing and
tighten the two holding screws (53.1).
If the performance of the instrument changes greatly after the replacement, the height
of W lamp should be adjusted (the horizontal position of W lamp is adjusted automati-
cally when the self-testing is done
upon switching on the power.). The
adjusting procedure is as follows:
a) Remove the halogen lamp holder.
b) Loosen the three screws (53.2) on
the halogen lamp holder with a watch
screwdriver.
c) Adjust the height of W lamp.
d) Tighten the three screws (53.2).
e) Insert the halogen lamp holder
back to the socket.
Replacement of the D2 lamp
The D2 lamp is plugged in the D2 lamp socket.
Open the light source housing cover. Remove the D2 lamp (53.D).
Replace it by a new D2 lamp. Clean the surface of D2 lamp with a clean cloth dampe-
ned with alcohol.
Insert the D2 lamp into its socket.
Note:
The light beam window should be facing the reflective mirror (53.G).
ATTENTION!
Both lamps and lamp holder are considerably heated during operation. Therefore, be sure
to turn off the power switch so as to cool down the lamp sufficiently before replacing it
The reflective mirror (53.G) should be covered with a paper bag to avoid being conta-
minated by fingers.
AC POWER 120VA
FUSE TYPE F
AC (100-130) 2A
AC (200-240) 1A
MODEL:
DATE: 2003 6
NO.: 03 - 1880 - 01 - 006
POW: AC220V 150VA 50 - 60 HZ
RS232
Printer
G
W
D
1
2
Picture 53 - Lamps changeover
Revisión 1 Junio-06 Manual de instrucciones 5448X000 Pág. 39
ANEXO I: CERTIFICADO CE
AUXILAB S.L.
Están diseñados y fabricados de acuerdo a:
Directiva 89/392/CEE, incluidas las modificaciones de la misma, y las regla-
mentaciones nacionales que la trasponen.
Directiva 89/336/CEE modificada sobre compatibilidad electromagnética.
Directiva 73/23/CEE modificada sobre seguridad eléctrica.
Y que se han aplicado las siguientes normas armonizadas (o parte de ellas):
UNE 292-1/-2/-2/A1, UNE 1050, UNE 547, UNE 953, UNE 294, UNE 894-1,
UNE 894-2, 954-1, UNE 60204-1, UNE 61010
Polígono Morea Norte, 8 31191 Beriain (Navarra) - Spain. Tel. 948 310 513 Fax 948 312 071
Internet: www.auxilab.es · Email: [email protected]
DECLARACIÓN CE DE CONFORMIDAD ESPECTROFOTÓMETROS ZUZI
SERIES 4418, 4419, 4488 Y 4481 de Auxilab, S.L a la Directiva de Máquinas
(89/392/CEE modificada) y a las reglamentaciones adoptadas para su transposición
NOMBRE DEL FABRICANTE / IMPORTADOR:
AUXILAB, S.L.
DIRECCIÓN:
Polígono Morea Norte, 8
31191 Beriáin (Navarra)
DECLARAMOS QUE:
ESPECTROFOTÓMETROS ZUZI
SERIES 4418, 4419, 4488 Y 4481
BERIAIN a 31 de Junio de 2004
Fdo: ALFONSO AINCIBURU SANZ
DIRECTOR/GERENTE
ENGLISH