Fault messages / troubleshooting /
disposal
16
Page 26
Error message Water
level low
first warning
If this symbol appears on the left-hand side of the bottom screen bar, the water level
in the device is getting low. Please top up with distilled or deionised water as
described in chapter 5.
second warning
This symbol appears after 24 operating hours if the device has not been topped up
with water after the first warning. Please top up with distilled or deionised water as
described in chapter 5.
The device can no longer be used safely if it contains less than 900 ml of water, and
the error message displayed appears. Please top up with distilled or deionised water
as described in chapter 5.
Error message
Handpiece not
connected
Handpiece not (or not correctly) connected.
Check whether the handpiece is correctly connected. The plug
must be fully inserted (this is indicated audibly by two clicks).
Error message
Handpiece sensor
The temperature sensor of the handpiece is no longer working properly. Please
replace the handpiece.
If two handpieces are being used to administer treatment, the user can continue to
administer treatment using the intact handpiece.