Device set-up
6.2 Starting up the device
6
Page 9
Distilled water must run through the whole system when installing the device for the
first time. Do this by starting up the device for 30 seconds (without selecting a
treatment) and then fill up again to the red mark.
Caution
!
This procedure must be repeated for each handpiece when it is connected for the
first time.
Switch device on
Switch on the device using the power switch (6).
Carry out a function test (see chapter 12).
Note:
You should always perform the following checks before using the device:
•
Are the housings and handpieces clean and disinfected?
•
Are the device's castors securely placed on the floor?
•
Is the power cable connected to an earthed socket?
Note:
When starting up the device, the handpieces should be connected and in their
mountings.
Switch device off
The device is switched off using the power switch (6).
To fully (all poles) disconnect the device from the mains, the power cable must be
disconnected.
Caution
!
We recommend removing water from the device via the drain tap if the device is not
being used for a prolonged period or is exposed to lower temperatures (e.g. in
transit).