Application information
3
Page 3
General instructions
for use
ZLipo
Med
has been designed to comply with international safety standards.
Prior to using ZLipo
Med
on a patient, the user/technician should become familiar with
the instructions for use and individual treatment methods to be used as well as the
indications, areas of application / contraindications, warnings and application
information. Additional sources of information about the treatment should also be
followed.
The device may only be used by doctors with appropriate specialist knowledge of fat
reduction. Make sure that all users are familiar with the use of the device and know
how to switch off the device immediately in an emergency.
Caution
!
During use, the device is to be located in a position allowing direct access to the
device´s central mains supply so that it can be disconnected from the mains at any
time.
Access to the treatment room must be restricted to staff who have been instructed in
the safety procedures.
Caution
!
Do not use the device if it has been dropped or exposed to other massive forces. Try
to minimise shocks and vibrations.
Caution
!
Make sure that the handpieces do not come into contact with hard materials that
could damage them.
Caution
!
Before use, ensure that the device is powered via a properly grounded mains socket
(e.g. electrical installation according to DIN VDE 0100 Part 710). The device must
only be operated with the supplied power cable. The power cable must be protected
against mechanical stress.
Caution
!
ZLipo
Med
is not suitable for use in areas with an explosive, flammable or combustive
atmosphere. Corrosive materials can damage the device. Dust and hairs should be
minimised.
Caution
!
To avoid the risk of electric shock, the plug must be disconnected from the power
supply before performing any cleaning or maintenance activities. Do not touch the
plug with wet hands.
Caution
!
Only accessories provided by Zimmer MedizinSysteme GmbH may be used.
Caution
!
In order to ensure adequate ventilation, the device must be kept at least 0.5 metres
away from other devices or walls.