Safety and maintenance
14.1 Safety
14
Page 23
This chapter describes general safety rules for the use of ZLipo
Med
, focusing in
particular on low-temperature and electrical safety. Suitably qualified users can
operate the device safely if it is correctly used and maintained. All staff who work
with or maintain the device must be familiar with the safety rules set out in the
instructions for use.
The safety of the person receiving treatment, the user and other staff have top
priority. The safety of the person receiving treatment is guaranteed primarily by well-
trained staff and a properly equipped treatment room. It is important that information
be provided to the person receiving treatment, and this must also include the nature
of the treatment.
ZLipo
Med
generates high voltages. Some components can discharge energy when
the device is switched off. For this reason, no parts of the housing may be removed.
Removing the housing can be dangerous.
Make sure that you always switch off the device before connecting or disconnecting
a handpiece.
Never use the device in the presence of flammable or combustible materials.
Water and electricity are a hazardous combination. Do not use ZLipo
Med
in a wet
environment. Do not immerse any components directly in liquid.
The use of third-party cables and accessories can lead to increased emissions or a
reduction in safety. Third-party accessories must not be connected to the device.
Do not use damaged cables.
As the manufacturer, Zimmer MedizinSysteme GmbH can only consider itself to be
responsible for the safety and reliability of the device if
•
the device is operated using a proper power outlet with earth contact and the
electrical installation complies with DIN VDE 0100 part 710,
•
the equipment is operated in accordance with the instructions for use,
•
extensions, readjustments or modifications are carried out only by persons
authorised by Zimmer MedizinSysteme,
•
the user has confirmed the functional safety and the proper operating condition
of the device before use,
•
the device is operated only by properly trained personnel
•
the device is not operated in hazardous areas and / or a combustive
atmosphere,
•
the device is immediately disconnected from the mains when penetrated by
liquid.
The device does not contain any parts that can be repaired by the operator.