Device set-up
6.2 Starting up the device
6
Page 8
Connect power cable Connect the power cable to the provided port (6) on the device and connect the
cable to the mains.
Note:
The device may only be connected to power outlets with a protective contact
Note:
All cables must be protected from pinching or other mechanical damage.
Connect applicators
Place the applicator in its mounting. Then connect it with the device by inserting the
plug into the correct socket until you hear two clicks.
To remove the applicator, press the buttons on either side of the plug
simultaneously.
Switch off the device and disconnect the power cable from the mains before
changing handpieces.
Caution
!
Failure to connect the applicators with the device correctly can lead to water spillage.
Fill with distilled water
Caution
!
Caution
!
Press the black slider to the right to open the doors on the rear of the device.
Open the filler neck by turning the white “Fill” cap a quarter-turn to the left and
removing it.
Connect the filler tube to the filler neck and slowly pour in distilled water until the red
marker is reached.
Only use distilled or deionised water, not tap water.
The water should be replaced every 3 months.