Device set-up
6.1 Site preparation
6
Page 7
Note:
Do not expose the device to direct sunlight. Operate the device at room
temperatures of between 10 and 26°C and air humidity of between 30 and 75%.
Do not set the device up close to fan heaters or air conditioners.
Note:
Make sure that the power switch of the device is set to “0”.
Site preparation
When setting up the device, the operator should take account of the spatial and
electrical requirements and transport and storage conditions. The following
requirements must be met before unpacking.
The premises must be supplied with adequate ventilation and a free flow of air. In
order to ensure optimum ventilation, the housing sides should be at least 0.5 metres
away from the wall or other elements that could block the flow of air.
The device can affect computers or other electrical or medical devices. A distance of
at least 2 metres from such devices should be kept.
The device must be connected to a power outlet with a protective contact. The
power supply data on the identification plate must be respected. The device should
not share its power supply with high-draw devices such as air conditioners. Ideally it
should be connected to a separate power supply with a separate protective switch.
The device must be operated in a corrosion-free atmosphere. Corrosive materials
such as acids can damage the electrical wiring, electrical components and the
surface of the treatment components.
Minimise dust and hair particles.
Do not place the device close to heaters or other devices that affect the temperature.
Safe operation depends on good condition.