Care and Maintenance
000000-1253-546 VISUCAM C 31.07.2003
60
Troubleshooting table
Fault Possible
cause
Remedy
Power switch not switched ON
Switch power switch to ON
No electric function
Power cable of basic unit not
connected
Establish connection to power outlet
No image on monitor
Power cable of monitor not connected
Establish connection to power outlet
Message ”Cable not
connected” on monitor
Signal cable of monitor not connected
to basic unit
Connect monitor signal cable to basic
unit
Lamp voltage control adjusted to
Minimum
Turn control clockwise until lamp
brightness is sufficient
No observation light
Halogen lamp faulty
Replace halogen lamp as described in
Section Replacing Lamps on page 56
Observation brightness not
continuously adjustable
Lamp control faulty
Call technical service
Fixation light is dark
Blinking diode faulty
Replace blinking diode as described in
section Replacing Lamps on page 57