13.3 Performance Verification Check
200 / 246
2660021169042 Rev. A 2018-03
13.3.3 Pass/Fail Conditions
Conditio
n
Description
Explanation
Image
Pass
If both scan line indicators pass partially or wholly
through the center of the cross-hairs on both LSO
and OCT Fundus images, the Test passes.
Both lines centered
Horizontal line
marginal. but OK
Vertical line
marginal, but OK.
Fail
If the cross-hair target shifts away from the scan
indicators when moving between OCT Fundus and
LSO images, the system fails the check.
Horizontal line
clearly within black
Vertical line clearly
within black