8 Operation
Instructions for Use
2660021169042 Rev. A 2018-03
8.5 Acquire Data (Scan)
94 / 246
2660021169042 Rev. A 2018-03
1
2
3
4
7
5
6
ð
The chinrest moves into place for the selected scan.
3. To change the
FOV
, click the
FOV (2)
appropriate for the
patient.
4. Left-click on the center of the
pupil (3)
in the
viewport (4)
.
ð
The pupil target appears in the viewport.
5. If needed, move the
chinrest settings (5)
to position the
image in the
viewport (4)
.
6. Click
Auto Focus (6)
and ensure that the
pupil target (3)
remains centered.
NOTE
If the fundus image becomes dark after you click Auto Focus,
recenter the pupil target and adjust the image brightness.
7. Click
Optimize (7)
.