1 Safety and Certifications
Instructions for Use
2660021169042 Rev. A 2018-03
1.4 Electromagnetic Compatibility (EMC)
24 / 246
2660021169042 Rev. A 2018-03
Voltage dips, short inter-
ruptions, and voltage
variations on power supply
input lines. IEC 61000–4–
11
<5% UT (>95% dip
in UT) for 0,5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (95% dip in
UT) for 5 sec
<5% UT (>95% dip
in UT) for 0,5
cycle 40% UT (60%
dip in UT) for 5
cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (95% dip in
UT) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the PLEX® Elite 9000 requires
continued operation during
power mains interruptions, it is
recommended that the PLEX®
Elite 9000 be powered from an
uninterruptible source.
Power frequency (50/60
Hz) magnetic field IEC
61000–4–8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
UT is the a.c. mains voltage prior to application of the test level.
Содержание PLEX Elite 9000
Страница 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Страница 54: ...Empty page for your notes...
Страница 106: ...Empty page for your notes...
Страница 174: ...Empty page for your notes...
Страница 182: ...Empty page for your notes...
Страница 210: ...Empty page for your notes...
Страница 218: ...Empty page for your notes...
Страница 224: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 228: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 233: ...Instructions for Use Index 2660021169042 Rev A 2018 03 2660021169042 Rev A 2018 03 233 246 Index...
Страница 245: ...Empty page for your notes...