Instructions for Use
13 Maintenance and Repair
2660021169042 Rev. A 2018-03
13.3 Performance Verification Check
2660021169042 Rev. A 2018-03
191 / 246
5. Select
Optimize.
6. Select
Cube 12 mm x 12 mm (512 x 512)
.
Figure 42: Acquire Screen showing a default Cube scan
7. Click
Auto Focus
to get a clearer image of the cross–hair test
pattern. (Use the focus arrows if your system does not have
Auto Focus
activated.) Besides focus, other adjustments
usually are not necessary, although possible.
8. Click
Capture
and then select OD or OS in the Select Eye dialog
that appears. The Review screen appears automatically.
9.
NOTE! Image appearance or signal strength have no
bearing on the co-alignment of the scan and fundus
images, which is what this test evaluates. If necessary,
you can adjust the brightness and contrast later in the
Analyze screen when evaluating the test, but otherwise
no adjustment is necessary.
Click
Save
and then either
Finish
or
ID Patient
to exit data acquisition. You will return to
the ID Patient screen.
10. Select the Performance Verification patient again and click
Analyze
.
11. In the Analyze screen, select the scan you just saved.
12. Select
En Face Report
in the right-hand column.
Содержание PLEX Elite 9000
Страница 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Страница 54: ...Empty page for your notes...
Страница 106: ...Empty page for your notes...
Страница 174: ...Empty page for your notes...
Страница 182: ...Empty page for your notes...
Страница 210: ...Empty page for your notes...
Страница 218: ...Empty page for your notes...
Страница 224: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 228: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 233: ...Instructions for Use Index 2660021169042 Rev A 2018 03 2660021169042 Rev A 2018 03 233 246 Index...
Страница 245: ...Empty page for your notes...