8 Operation
Instructions for Use
2660021169042 Rev. A 2018-03
8.6 Check Scan Quality
102 / 246
2660021169042 Rev. A 2018-03
6. Scan is complete in all windows (no missing data).
7. Light intensity is uniform across the image.
8. Scan passes: RPE Acceptance Criteria [
}
103]
.
Retina is not too low in the scan, which impacts sub-RPE slice
detection.
9. Scan passes: Signal Quality Acceptance Criteria [
}
103]
.
Signal strength is 6 or higher; shadows and dark spots exhibit
floaters or disease.
10. Scan passes: Decorrelation Tails Acceptance Criteria [
}
104]
.
Scan shows accurate motion, no decorrelation tails, no vascu-
lature in the RPE layer, and superficial and deep retinal layers
look appropriately different.
11. Scan passes: Segmentation Acceptance Criteria [
}
104]
.
Presence or absence of flow appears in layers as appropriate.
Result
ü
If any criteria listed above are not met, retake the scan.
8.6.4 Acceptance Criteria
When reviewing PLEX® Elite 9000 Angiography Scans for accept-
ability, consider the following:
• RPE Acceptance Criteria [
}
103]
• Signal Quality Acceptance Criteria [
}
103]
• Decorrelation Tails Acceptance Criteria [
}
104]
• Segmentation Acceptance Criteria [
}
104]
Consider all these possibilities before accepting OCT Angiography
scans for further analysis.
Содержание PLEX Elite 9000
Страница 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Страница 54: ...Empty page for your notes...
Страница 106: ...Empty page for your notes...
Страница 174: ...Empty page for your notes...
Страница 182: ...Empty page for your notes...
Страница 210: ...Empty page for your notes...
Страница 218: ...Empty page for your notes...
Страница 224: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 228: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 233: ...Instructions for Use Index 2660021169042 Rev A 2018 03 2660021169042 Rev A 2018 03 233 246 Index...
Страница 245: ...Empty page for your notes...