PK=aÉëÅêáéíáçå
léíáçåë
3.4.8.2. SESI detector
The Secondary Electrons Secondary Ions detector (SESI detector) is suitable to detect secondary
electrons as well as secondary ions. The SESI detector replaces the SE2 detector.
Depending on the polarity of the collector voltage either electrons or ions scattered from the spec-
imen (
5
) are attracted by a collector grid (
4
) and accelerated to the converter. In the converter,
both electrons and ions are converted into secondary electrons which are used to generate an
image.
The SESI detector can be operated in two modes: SE mode and Ion mode.
1
FIB column
4
Collector grid
2
GEMINI
®
column
5
Specimen
3
SESI detector
Operating mode
FIB mode
Detected signals
Typical application
SE mode
(typical collector
v300 V)
SEM
Secondary electrons
Topography
FIB
Ion mode
(typical collector
voltage -4 kV)
-
Secondary ions
Crystal orientation contrast, material con-
trast e.g. imaging of corrosion/oxidation
processes in metals
FIB
Fig. 3.14: Schematics of the SESI detector
Содержание Crossbeam 340
Страница 1: ...Crossbeam 340 Crossbeam workstation Instruction Manual ...
Страница 24: ...OQ çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF OK p ÑÉíó p ÑÉíó ÉèìáéãÉåí ...
Страница 60: ...SM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF RK fåëí ää íáçå ...
Страница 120: ...NOM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Страница 126: ...NOS çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NNK ÄÄêÉîá íáçåë ...
Страница 130: ...NPM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...
Страница 133: ......