zebris Medical GmbH
FDM-T Technical Data and Operating Instructions
Page 48/65
6.3.2 Connection of the System
For connecting the treadmill to the power supply, please additionally observe the respec-
tive instructions in the Operating Instructions given by the treadmill manufacturer.
WARNING
The connection for the treadmill and FDM-T power supply unit must
be plugged directly into the wall socket. It is not permissible to use
extension cables and/or multiple sockets.
WARNING
For connecting the different system components to the mains supply
it is absolutely essential to keep to the following illustration exactly,
in order to avoid the patient coming into contact with any dangerous
touch currents. For further information on this, please see Section
8.2 of this user manual. It is essential to observe the strict separation
between the power supply of the system components that can be
located within the vicinity of the patient (treadmill incl. platform) and
those that must be located outside the patient's vicinity (PC, video
camera, illumination).
WARNING
Set all connection cables in a way that prevents patients, or persons
taking part in the measuring procedure from trip over them or dam-
aging them mechanically. If necessary anchor the cables with adhe-
sive tape to the floor or use protection covers.