6.3
Connection of the measuring system to mains supply
6.3.1 Power supply of the FDM-T Sensors
For connecting the FDM-T sensor to the power supply, connect the power supply unit to
the mains socket and the power socket on the interface box.
WARNING
For operating the FDM-T sensor, exclusively use the power adapter
approved by zebris, which is suitable for the power supply for all the
FDM-T systems.
REF.-No.
181.1515 / PS Treadmill V2
Input
Output
Cable
Length
100 - 240 V AC
16V DC
Mains lead
1.7 m
50 - 60 Hz
60 W
DC lead
5 m
Pin arrangement / polarity
NOTE
Before connecting the measuring system to the power supply, com-
pare the label specifications on the power supply unit and on the
treadmill as regards the mains voltage and mains frequency, with
the local characteristics. Connect only if they agree.
WARNING
Before connecting or using the measuring system, carry out a visual
check of the power supply unit, power connection lead and socket,
as well as the ground contacts. Damaged power supply units, leads
or plug and socket devices are to be replaced immediately by an
authorized person.
„
-
“ Contact:
outer contact
of hollow plug
„
+
“ Contact:
inner contact
off hollow plug