zebris Medical GmbH
FDM-T Technical Data and Operating Instructions
Page 25/65
3.7.2 Synchronization Output (SYNC-OUT)
If a third party device is connected to the synchronization output SYNC-OUT then depend-ing
on the setting of the configuration window of the application software, a FDM-T sys-tem-
controlled, will trigger a synchronized measurement of the third party device either via
start/stop or “sample by sample” mode.
Electrical Specifications
Output Resistance
100
High-Level
≥ 2.0V
Low-Level
≤ 0.8V
Built in LEMO
– Socket (Interface-Box)
Series „00“, 4-Pin, Coding 0°
LEMO-Part-No.: EPG.00.304.NLN
View
Socket, Front Side
Socket Coding: 0°
Respective Plug for SYNC-OUT
LEMO-Part-No.: FGG.00 304.CLADxxxx
Pin-Assignment
Pin 1
+5V
Pin 2
GND
Pin 3
Activ_OUT
Pin 4
Clk_OUT
View
Plug, Solder Side
Plug Coding: 0°