
zebris Medical GmbH
FDM-T Technical Data and Operating Instructions
Page 24/65
3.7.1 Synchronization Input (SYNC-IN)
If a third party device is connected to the synchronization input SYNC-IN then depending
on the setting of the configuration window from the application software the measurement
will start/stop or “sample by sample” synchronized by a signal from the third party device.
Input is protected against faulty polarisation and pin 1 is set to +5V ("1") by an internal pull-
up- resistor (2.7 k
). If this input is set to 0 V ("0") i.e. by a switch or break contact than the
SYNC-IN is triggered.
Electrical Specifications
Input Resistance (Pull-Up 5V)
2.7k
V
IH
(High-Level Input Voltage)
≥ 2.0V
V
IL
(Low-Level Input Voltage)
≤ 0.8V
Required min. pulse time for triggering
1ms
Eingebaute LEMO
– Buchse in der Interface-Box
Serie „00“, 3-polig, Codierung 30°
LEMO-Teile-Nr.: EPA.00.303.NLN
View
Socket, Front Side
Socket Coding: 30°
Respective Plug for SYNC-IN:
LEMO-Part-No.: FGA.00 303.CLADxxxx
Pin Assignment
Pin 1
Clk_IN
Pin 2
Activ_IN
Pin 3
GND
View
Plug, Solder Side
Plug Coding: 30°