Wrong or Erratic Data:
1-
Check the sensor Properties screen for possible errors.
2-
Verify slopes, offsets, or custom functions.
3-
Check battery voltage and all connections.
SDI-12 Communications
No SDI-12 Response:
1-
Verify power to the System 5000
.
TM
2-
Check all SDI-12 connections, including DATA and GND.
3-
Check the pin out of the SDI-12 sensor for correct wiring.
4- Make sure all sensors have a unique address.
GOES Transmissions
GOES not transmitting:
1-
Check power to the GOES Radio, do so under a load (e.g. a transmission).
2-
See if the GOES time is correct, if it is stopped, it indicates power to the radio was lost,
stopping the clock. The radio will not transmit if the GOES clock is not running.
3-
Check all antenna wires.
4-
Check antenna alignment.
5-
Use a power meter to measure the power output of the radio and the reflected power.
System 5000
Maintenance and Troubleshooting 14-3
TM
Содержание WaterLog Series
Страница 1: ...Model System 5000TM Owner s Manual Revision 1 4 2...
Страница 2: ......
Страница 8: ......
Страница 24: ...2 10 Hardware Overview System 5000TM...
Страница 58: ...5 14 Inputs System 5000TM...
Страница 128: ...10 14 System Setup General Setup System 5000TM...
Страница 144: ...12 6 System Setup Inputs Outputs System 5000TM...
Страница 168: ...14 4 Maintenance and Troubleshooting System 5000TM...
Страница 170: ...A 2 System 5000 Specifications System 5000TM...
Страница 172: ...B 2 Pressure Sensor Option Module Specifications System 5000TM...
Страница 174: ...C 2 Analog Digital Option Module Specifications System 5000TM...